Byte
Name
Description
6
Test
Active
Command status
7
Stand-
still
Inside
ZSR
Center
Zero
Below
Zero
Over-
load
Above
Max
ADC
error
Device status
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
The gross value (hex:000004D2 <=> 1234) can be read from bytes 0
3.
When the "Overload", "Test Active" or "ADC error" bits are set, the read value is invalid.
Negative values are output in the second complement.
12.2.5
Special hints for DeviceNet and EtherNet-IP
With these fieldbus types, the sequence of the bytes (only applicable for words and
individual bytes) is inverted.
With long words, this problem does not arise due to compensation by the firmware.
Sequence of bytes 0
3, e.g. with device type and software version, see table below.
Standard sequence
Sequence for DeviceNet and EtherNet-IP
Byte 0
TYPE MSB
Byte 0
SUBVERSION
Byte 1
TYPE LSB
Byte 1
MAIN VERSION
Byte 2
MAIN VERSION
Byte 2
TYPE LSB
Byte 3
SUBVERSION
Byte3
TYPE MSB
Consequently, the sequence on the PLC side must be changed when using the DeviceNet
and EtherNet–IP field bus types.
12.2.6
Fieldbus register
12.2.6.1
Register 0: I/O status bits (read)
Dynamic status
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Input 3
Input 2
Input 1
Byte 1
Output 3
Output 2
Output 1
Byte 2
Limit 3
Limit 2
Limit 1
Byte 3
12.2.6.2
Register 1: Scale status (Read)
Dynamic status
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
OUT
STND
INZSR
CZERO
BELOW0
OVL
>Max
ADUERR
Byte 1
E9
E6
E1
E3
E7
Byte 2
PowerFail
ActionAc-
tive
CmdError
X3 Process Indicator PR 5410
12 Fieldbus interface
Minebea Intec
EN-300