16 Error messages
16.1 Error messages measuring circuit
The internal weighing electronics can generate error messages, which are shown on the
weight display.
Display
VNC text
Error and possible cause
Remote dis-
play
Error 1
Arith. error
Internal arithmetic overflow:
-
Faulty calibration values.
Error 1
Error 2
Overload
The measuring signal is higher than Max + (x d):
-
Wrong setting.
-
Too much weight on the scale.
Error 2
Error 3
Ext. meas. device error
Measuring input open:
-
The measuring signal is higher than the per-
missible range of 36 mV.
-
Measuring cable is interrupted (cable break
detection).
-
Other hardware defect.
Error 3
Error 4
Value exceeds display
The weight value is not displayed:
-
Too many digits have been set.
Error 4
Error 6
No sense voltage
No sense voltage:
-
Load cells not connected.
-
Sense line or supply line is interrupted.
-
Wrong polarity or sense voltage is low.
Error 6
Error 7
Negative input
Negative measuring signal:
-
Wrong polarity of load cell signal.
-
Wrong polarity of load cell supply voltage.
Error 7
Error 9
No values from scale
Internal weighing point:
The measuring signal is higher than the permis-
sible range of 36 mV. Cannot read weight values
from ADC (analog-digital converter).
-
Error in weighing electronics board.
-
Defective load cell.
-
Cable break.
Error 9
Negative signal Negative weight
Display area
Max
-¼ d
+¼ d
0
Overload none EOC
X3 Process Indicator PR 5410
16 Error messages
Minebea Intec
EN-324