16.2 Error messages for xBPI scales
Display
VNC text
Error and possible cause
Remote dis-
play
Error 3
Ext. meas. device error
No weight values can be read from scale:
-
Scale error
Error 3
Error 4
Value exceeds display
The weight value is not displayed:
-
Too many digits have been set.
Error 4
Error 9
No values from scale
No communication with xBPI scale:
-
Cable break.
-
Internal scale error.
-
The scale is not connected to the power sup-
ply.
Error 9
Error 11
No weight data
No weight display:
-
Another weighing point was selected.
Press
to assign the new weighing point to
the device.
Requirements:
- [Serial ports parameter]- [xBPI port]- [Slot
1/2 RS 485] is selected.
Error •
Error 12
Scale not ready
The scale is not ready for weighing:
-
The device is in the warm-up phase.
-
The device is in automatic taring mode.
-
The device was switched on with the scale
loaded.
Switch the device off and on again.
Error <
Error 15
Wrong serial number
Serial number of scale does not match the num-
ber set in the device.
Error?
16.3 Error messages for Pendeo load cells
Display
VNC text
Error and possible cause
Remote dis-
play
Error 2
Overload
The measuring signal is higher than Max + (x d):
-
Wrong setting.
-
Too much weight on the scale.
Error 2
Error 4
Value exceeds display
The weight value is not displayed:
-
Too many digits have been set.
Error 4
16 Error messages