Display
VNC text
Error and possible cause
Remote dis-
play
Error 8
Incompatible units
Incompatible units:
-
Incorrect calibration values e.g. due to incor-
rect calibration
Error 8
Error 9
No values from scale
No communication with Pendeo load cell:
-
Cable break.
-
Internal scale error.
-
The junction box is not connected to the sup-
ply voltage.
Error 9
Error 12
Scale not ready
The scale is not ready for weighing:
-
At least 1 load cell gives an error status or is
defective (no communication).
Error <
Error 13
Wrong configuration
The number of load cells does not match the con-
figuration.
Error =
Error 15
Wrong serial number
Serial number of scale does not match the num-
ber set in the device.
Error?
16.4 Error messages during calibration
Message
Possible causes
This message displays, if the maximum load is too high.
The maximum load of the scale can be increased retroactively. When the
capacity is reduced, however, a message is displayed when the new ma-
ximum load is lower than the CAL weight ([Calibrated at]).
This message is displayed if the selected resolution is too low, e.g., 5 kg.
X3 Process Indicator PR 5410
16 Error messages
Minebea Intec
EN-326