Message
Possible causes
This message displays, if the scale is not stable.
Remedy
-
Check the mechanical function of the scale.
-
Adapt the filter setting; reduce the resolution.
-
Adapt the standstill conditions.
This message is displayed if the weight on the scale is < the dead load
after input of the weight value.
The maximum capacity is not an integer multiple of the scale interval.
16.5 General error messages
Display
Error and Possible Cause
Err 21
BIOS is too old for the firmware.
Err 22
Firmware is too old for the BIOS.
Err 23
Boot not possible:
-
ROM error
-
No firmware found
Err 25
Plug-in card combination is not permissible.
Err 29
"Flash" process has failed.
Err 31
The digit group for IP address/mask is not within 0–255, or is invalid.
Err 32
The IP address is already in use.
Two instruments with identical IP addresses.
Err 33
Wrong parameter
Err 34
No Ethernet cable (LINK) found
Err 41
Major system error
Err 42
Major system error in controls.
Err 43
Monitoring time has expired.
Err 44
Cannot save to EAROM.
X3 Process Indicator PR 5410
16 Error messages
Minebea Intec
EN-328