Display
Error and Possible Cause
Err 86
xBPI weighing point is selected.
This weighing point can only be calibrated via VNC.
Err 87
Sequence no. was not found in alibi memory.
Err 88
Sequence no. cannot be changed because the alibi memory is activa-
ted.
Err 89
No EW-Com interface assigned.
Err 90
No printer interface assigned.
Err 91
Linearization is active.
Err 92
xBPI interface cannot be opened.
Err 93
Too many Pendeo load cells were found.
Err 94
Search for Pendeo load cells is defective.
Err 95
Communication error with Pendeo load cells.
Err 96
Not all Pendeo load cells are the same type.
Err 97
Dead load cannot be set for Pendeo load cells.
Err 98
Weighing point cannot be assigned to the Pendeo load cell.
Err 99
xBPI communication is not assigned.
16.6 Error numbers @ "LAST_ERROR"
16.6.1
Weighing point error
Number
Display
Cause
2
in use
Weighing point is in use, e. g., by an application.
6
test active
Test is active, no weights.
7
cali active
Calibration is active, no weights.
8
no standstill
No standstill of the scale.
13
tare is active
Tare is active.
15
weight is dimmed
Weight is not legal-for-trade (<0 or >max.).
16
weight has error
Weight error.
17
scale not ready
Scale is not ready.
18
cannot tare below zero
Taring below zero is not possible.
102
timeout
Only on xBPI scales: Time limit exceeded when sending a com-
mand to the scale.
142
calibration active
During calibration, taring and zeroing is not possible.
147
no zeroset
Zeroset outside of the zero setting range is not possible.
149
Busy
The scale is currently busy with another query.
255
hardware error
Weighing point is faulty.
X3 Process Indicator PR 5410
16 Error messages
Minebea Intec
EN-330