3. Connection
※
1 When tension is applied with the application of tension type or universal (compression
/tension) type of load cell, and display of “+” direction is required, connect “Green”
with Terminal No.4 and “Blue” with Terminal No.2 individually. As there is a case
which standard wiring color is different, please confirm the inspection data sheet of the
load cell being used.
※
2 When the length of CAB
−
502 is more than 30 m totally, there may have the case that
the accuracy is out of warranty because the resistance of cable makes the input voltage
of the instrument decreased.
※
3 When the length of CAB
−
502 is more than 10 m totally, or the zener barrier is used in
the system, CALIB is not applicable.
3
−
3
−
1. Connection with strain gage applied transducers
The instrument can connect with strain gage applied transducers, such as load cell, pressure
transducer and so on. Here, we will describe the example of connections with load cell, so the
connection with another type of strain gage applied transducers shall be proceeded in the same
way.
1
Connection with 1 piece of load cell and CSA
−
521
3 m
WHT
BL
U
※
1
YEL(Shield)
GRN
※
1
RED