Electronic calibration by CALIB input
Procedures
1
Set the load cell in the condition of the initial
load (tare weight).
When the initial load (tare weight) is exceeding
±
0.25 mV/V at the input conversion, cancel the
initial load (tare weight) referring to the
paragraph 4
−
3.
ZERO trimmer
SPAN trimmer
2
ZERO adjustment
Adjust the voltage output value to become at
0.000 V with ZERO trimmer.
2