19
6. Specifications
6
−
1. Specifications
・
Bridge power supply
DC10 V
±
0.3 V
・
Applicable transducers
1 pieces of strain gage applied transducer(350
Ω
) is connectable.
・
Input range
0.5 mV/V to 1.5 mV/V
・
Sensitivity
10 V output at the input of 0.5 mV/V
・
Sensitivity adjustment range2 000 times
adjustable from 1/1 to 1/4
・
Output voltage
±
10 V
・
Output load resistance
2 k
Ω
or more
・
Zero adjustment range
Adjustable
±
0.25 mV/V by trimmer
※
Adjustable the tare weight compensation by mounting the
resistance.
・
Non
−
linearity
0.02 %F.S.
・
Temperature coefficient
Zero point
±
1
μ
V/
℃
(Input conversion)
Sensitivity
±
0.01 %F.S./
℃
・
CALIB
0.5 mV/V
±
0.001 mV/V
・
Frequency response range
1 kHz(
−
3 dB
±
1 dB)
6
−
2. General specifications
・
Operating temperature/humidity range
Temperature
−
10
℃
to 50
℃
Humidity
Less than 85 %RH (Non condensing.)
・
Power supply voltage
DC24 V
100 mA
・
Outline dimensions(W
×
H
×
D)
80 mm
×
70 mm
×
20 mm
・
Weight
Approx. 50 g
・
Terminals
236
−
110(WAG0)
:
CN1
6
−
3. Standard specifications at the time of shipment
・
Sensitivity
10 V output at the input of 1 mV/V
6
−
4. Accessories
・
Instruction manual
1 piece
・
Minus driver(small)
1 piece
Summary of Contents for CSA-521
Page 1: ...Transmitter CSA 521 Instruction Manual EN294 1145 G ...
Page 2: ......
Page 10: ......
Page 31: ......