2. Calibration procedure

The explanation of actual load calibration procedures for this unit are two as follows:

1

  

CSA

591

 

calibration

 procedure 

by

 

the

 

actual

 

load

 

 

5

2

1

2

  

CSA

591

−99 

(with

 

automatically

 

zero

 

function)

 calibration procedure 

by

 

the

 

actual

 

load.

 

5

2

2.

 

5

2

1.

 CSA-591 

Calibration

 

procedure

 

by

 

the

 

actual

 

load

.

Procedures

1

Set the gage onto tie bar, etc.
Connect the digital volt meter, tester, etc.,
between “+” and “

” of the check pin, or

between “+” and “

” of the voltage output at

rear panel.
Set the gage to the condition of zero adjustment.
Set the rotary switch for zero adjustment to the
position of zero.

“ZERO” adjustment
polarity switch
“ZERO” adjustment
rotary switch

The front panel was removed.

2

Zero adjustment
Confirm whether the output voltage can be
adjustable to

±

0.2 V or less by turning corse

adjustment trimmer.

1

Adjustable
Adjust the output voltage to approx.

±

0.2 V or

less by “ZERO” coarse adjustment trimmer,
and then adjust the output voltage to 0.000 V
by “ZERO” fine adjustment trimmer.

2

Not adjustable
Set the “ZERO” adjustment rotary switch at
the position in which the output voltage
becomes

±

10 V or less.

When the output voltage does not change, set
the “ZERO” adjustment polarity switch to the
position which the output voltage becomes

±

10

V or less. After that, set the output voltage to
approx.

±

0.2 V or less by “ZERO” coarse

adjustment trimmer, and adjust the output
voltage to 0.000 V by “ZERO” fine adjustment
trimmer.

When both LED1 and LED2 is turned off, the
output voltage outputs approx.

が±

50 mV or

less.

When LED1 lights on, the output voltage is
approx. +50 mV or more. When LED2 lights on,
the output voltage is approx.

50 mV or more.

rotary switch

“ZERO” coarse
adjustment trimmer

“ZERO” fine
adjustment trimmer
“GAIN” rotary switch

“GAIN” trimmer

CAL switch

LED1

Check pin “+”

Check pin “

LED2

V

OUT “

V

OUT “

Rear panel

3

Sensitivity adjustment

1

Take the rated load to the gage, and turn the
“GAIN” rotary switch so that the output
voltage may approach 10.00 V most.

2

Adjust the output voltage to 10.000 V by using
the “GAIN” trimmer.

The sensitivity adjustment range is

±

10 V of

output voltage at the input of

±

300

μ

ST to

±

2000

μ

ST.

Summary of Contents for CSA-591

Page 1: ...Transmitter CSA 591 Instruction Manual EN294 1141 F ...

Page 2: ......

Page 3: ...ains installation procedures and connecting method and also operating method for the Transmitter CSA 591 Make use of it properly after reading through the manual carefully Be sure to deliver the manual to the end user Moreover the end user should keep the manual at hand after reading it over ...

Page 4: ...t do this Take care and For reference Be sure to read these items where these marks are attached Warning Warning may cause injury or accident that may harm to the operator Don t do these things described here Caution during operation and working Be sure to read the item to prevent malfunction Mark during operation Press the switch ...

Page 5: ... and or high temperature area Don t use the instrument in a high humid area Don t install the instrument where there are vibrations and shocks Don t use the instrument where there is excess of dusts and fine particles Don t use the instrument where there are corrosive gas and salt and like that Don t install the instrument where there is rapid change of temperature and humidity Don t install the i...

Page 6: ...strument install as referring to the following figures and secure the space around the instrument Each dimensions of the instrument and required dimensions for the environmental spaces are as follows Outline dimensions Panel cut size Unit mm ...

Page 7: ...e be sure to make calibration If calibration will not be made the correct measuring results may not be obtained nor which may cause malfunction in the instrument and there may exist damage in peripheral equipments Besides even though calibration has been made there may occur the similar case when the results are not correct so make calibration again Warning In case of using the instrument check th...

Page 8: ...ny name in the cover page Delete the company name in the contents Mar 2018 DRW No EN294 1141 F Due to ECN FN18 02019 Correction INDEX P10 5 2 Calibration procedure CSA 591 calibration by the actual load CSA 591 calibration procedure by the actual load CSA 591 with automatically zero function by the actual load CSA 591 99 with automatically zero function calibration procedure by the actual load 5 2...

Page 9: ... 6 4 3 Connection 7 4 3 1 Connection of power supply and ground 8 5 Calibration procedure 9 5 1 Preparation 9 5 2 Calibration procedure 10 10 5 2 1 CSA 591 Calibration procedure by the actual load 12 5 2 2 CSA 591 99 Calibration procedure by the actual load 6 Options 14 6 1 Automatic zero CSA591 P99 14 6 1 1 Equivalent circuit of input output section 14 6 1 2 Function 14 7 Trouble shooting 15 7 1 ...

Page 10: ......

Page 11: ... zero corse adjustment 3 ZERO coarse adjustment trimmer It is a trimmer for zero coarse adjustment 4 ZERO fine adjustment trimmer It is a trimmer for zero fine adjustment 5 GAIN coarse adjustment rotary switch It is a rotary switch for sensitivity coarse adjustment 6 GAIN fine adjustment trimmer It is a trimmer for sensitivity fine adjustment 7 CAL switch CAL value is output when switch is turned ...

Page 12: ...from factory Don t turn it 12 Switch of ON for automatically zero A Z Automatically zero operates by turning ON 13 Switch of OFF for automatically zero The voltage in which the voltage of PC board for automatically zero is subtracted by turning OFF is output It is used when the initial zero is adjusted 14 LED It lights on while the automatically zero is working 2 2 Rear panel 1 Terminal block It i...

Page 13: ...e area Don t use the instrument in a high humid area Don t install the instrument where there are vibrations and shocks Don t use the instrument where there is excess of dusts and fine particles Don t use the instrument where there are corrosive gas and salt and like that Don t install the instrument where there is rapid change of temperature and humidity Don t install the instrument near the devi...

Page 14: ...nstrument install as referring to the following figures and secure the space around the instrument Each dimensions of the instrument and required dimensions for the environmental spaces are as follows Outline dimensions Panel cut size Unit mm ...

Page 15: ...tput 7 DC24 V Power supply input terminal DC24 V is connected 8 DC24 V Power supply input terminal DC24 V is connected 9 C Input 1 connecting terminal RED 10 B Output 1 1 connecting terminal BLU Tie bar gage is connected 11 B Output 1 2 connecting terminal WHT g g 12 GND Shield Connected internally with terminal No 4 Shield of gage is connected D class single grounding 13 BUSY Output terminal for ...

Page 16: ...own or not to apply strong impact The tightening torque of the screw on the terminal block is 0 8 N m at the maximum Crimp terminal that suits the terminal block of this unit is shown in the table below Width of Crimp terminal Suitable crimp terminal 6 0 mm or less 1 25 3 or Y type 1 25 3 5 Connecting cable with the instrument should be away from the noise source such as power supply line and or I...

Page 17: ...13 15 COM External A Z DC24V V OUT Open collector output Rear panel GN RD WH GN WH YL Shield BK BK WH RD D type single grounding FA409 404 Connected by connector UX010 LB UX010 LB or Warning Connection with external control input should be made securely according to the figures If neglected it may cause an unexpected failure and or malfunction to the instrument For the connection with external con...

Page 18: ...ning Connections with voltage outputs should be made securely according to the figures and also within specified load resistance If neglected it may cause an unexpected failure and or malfunction to the instrument For the connections with voltage outputs be sure to apply shielded cable and the shielded cable should be connected with the F G terminal of the instrument If not connected it may cause ...

Page 19: ...t is not correct So make precise calibration again Warning When you connect the tester and digital volt meter etc please connect them correctly before turning on the power supply 5 1 Preparation 1 According to the 4 Connection method connect this unit and a tie bar gage correctly and turn on the power 2 In case of CSA 591 99 with automatically zero function begin the adjustment after set the condi...

Page 20: ... 000 V by ZERO fine adjustment trimmer 2 Not adjustable Set the ZERO adjustment rotary switch at the position in which the output voltage becomes 10 V or less When the output voltage does not change set the ZERO adjustment polarity switch to the position which the output voltage becomes 10 V or less After that set the output voltage to approx 0 2 V or less by ZERO coarse adjustment trimmer and adj...

Page 21: ...step 3 ZERO adjustment polarity switch ZERO adjustment rotary switch ZERO coarse adjustment trimmer ZERO fine adjustment trimmer GAIN rotary switch GAIN trimmer CAL switch Check pin The front panel was removed 5 Confirm the CAL value When the transmitter is replaced it is possible to adjust the sensibility of the replaced transmitter in almost the same state of the calibration when CAL switch is t...

Page 22: ...s 10 V or less After that set the output voltage to approx 0 2 V or less by ZERO coarse adjustment trimmer and adjust the output voltage to 0 000 V by ZERO fine adjustment trimmer When both LED1 and LED2 are turned off the output voltage outputs approx 50 mV or less When LED1 lights on the output voltage is approx 50 mV or more When LED2 lights on the output voltage is approx 50 mV or more ZERO co...

Page 23: ... when CAL switch is turned on and the output voltage is recorded 6 Set the automatic zero off switch to ON side Turn on the automatic zero on switch or external automatic zero Calibration is completed Any difference of automatic zero is changed depending on the input condition noise etc Automatic zero off switch VR1 LED1 A Z OFF A Z OFF ON Automatic zero on switch PCB for auto zero ...

Page 24: ...t of input output section DC24 V DC5 V 10 k ohm A Z COM COM BUSY DC5 V 0 022 μST 6 1 2 Function 1 Automatic zero ON Automatic zero function works when the connection between A Z and COM is shorten for about 50 ms or more and the voltage output becomes zero after about 500 ms However the difference of automatic zero is changed depending on the input condition external noise etc The operation of aut...

Page 25: ...strument check by the following procedures However when you can t find any applicable item nor solve the symptom of trouble even after you have taken some measures contact us OK NG Supply power again Operation Normal Operation Start measurement Abnormal operation Execute trouble shooting ...

Page 26: ... and situation at site in details 1 NO YES YES YES NO NO Voltage output doesn t operate Load resistance is 2 kΩ or more Set load resistance over 2 kΩ Automatic zero function External control input does not work NO YES Connecting method is OK YES NO Connect securely according to the paragraph 4 3 Automatic zero operation is within the capable range YES NO Set automatic zero within the operation ran...

Page 27: ...tput the voltage Voltage output fluctuates abnormally Zero adjustment of voltage output cannot be made Inform us about the contents of failure and situation at site in details Proceed zero adjustment referring to the paragraph 5 2 Confirm the condition of connection and grounding ...

Page 28: ... by trimmer Non linearity 0 1 F S Temperature coefficient Zero point 10 μV Input conversion Sensitivity 0 05 F S CALIB 300 micro ST 15 micro ST 1 point Frequency response range Approx 100 Hz or approx 30 Hz Selectable by dip switch on PC board 8 2 General specifications Operating temperature humidity range Temperature 0 to 50 Humidity Less than 80 RH Non condensing Power supply voltage DC24 V DC18...

Page 29: ... 500 ms Accuracy of automatic zero within 20 mV V Battery backup time Approx 10 years in the room temperature Lithium electric cell battery use External automatic zero input 1point BUSY output 1 point output during the operation of automatic zero Open collector output 8 6 Outline dimensions Panel cut size Unit mm ...

Page 30: ...gent from which you have purchased 9 2 Repair Before asking repairs make checks once again that the connection setting and adjustment for the instrument have finished properly by referring to 9 Trouble shooting Especially make checks whether the connections of gages are disconnected or cut off After that still there may be found some defects in the instrument contact our sales office or sales agen...

Page 31: ......

Page 32: ...e Product Sales Management 1 1 1 Katase Fujisawa shi Kanagawa ken 251 8531 Japan Tel 81 466 23 2681 Fax 81 466 22 7191 Sensing Device Business Unit FUJISAWA PLANT 1 1 1 Katase Fujisawa shi Kanagawa ken 251 8531 Japan Tel 81 466 22 7151 Fax 81 466 22 1701 KARUIZAWA PLANT 4106 73 Miyota Miyota machi Kitasaku gun Nagano ken 389 0293 Japan Tel 81 267 31 1309 Fax 81 267 31 1353 HOMEPAGE ADDRESS http ww...

Reviews: