7-1.
Execute trouble shooting
Execute trouble shooting.
Does supply power voltage
satisfy the specifications?
Supply power source suitable for
the specifications.
Inform
us
about
the
contents
of
failure
and
situation
at
site
in
details.
1
NO
YES
YES
YES
NO
NO
Voltage output doesn’t
operate.
Load resistance is
2 k
Ω
or more.
Set load
resistance
over 2 k
Ω
Automatic zero
function (External control
input) does not work
NO
YES
Connecting method is OK?
YES
NO
Connect securely
according to the
paragraph 4
−
3.
Automatic zero
operation is within the
capable range
YES
NO
Set automatic
zero within the
operation range.