90
Set the connection range to DC
・
mV for the measuring instrument such as tester
and so on.
①
Remove the connecting cable from the terminals on the strain gage applied transducer.
②
Measure the voltage between terminal A and terminal C.
①
Short the terminal B, D and A, G.
②
According to the paragraph 7
−
15, set the load display section to the
monitoring condition for strain gage applied transducer.
The display stabilizes
at an optional value.
8
Inform
us
about
the
contents
of
failure
and
situation
at
site
in
details.
Inform
us
about
the
contents
of
failure
and
situation
at
site
in
details.
Start of
measurement
Same condition
YES
NO
NO
NO
YES
YES
The voltage between A
and C is just like the
setting of SW1, and
stable as well.
According to the procedures in the chapter 4, check on various kinds
of connecting conditions. Especially, check that if there may have effects
from the noise source such as invertor or like that.