6. Operating procedures
We’ll show the operating procedures for the instrument on each key on the front panel.
Warning
●
Operation with each key should be made after interrupting the
measurement. If it is made during measurement, it may cause an
unexpected malfunction.
●
Key operation in Measurement mode can be effective by pressing it
for about one second or so.
6
−
1. “A/Z
A
” key
6
−
1
−
1. Operations in Measurement mode
At the same time the Tare weight cancellation
(
A/Z
)
function activates and load display value
changes into the net weight display value, “A/Z” on the status display will light up and load
display will become “0”.
6
−
1
−
2. Operations in another modes
(1) Carrying set value
By pressing the “A/Z
A
” key with the condition of displaying various kinds of set values, the
flashing digits for set value will carry from 10
0
, 10
1
, 10
2
, 10
3
to 10
4
in order.
(
However, the range of carry differs depending on the number of digits of set values and yes/no
of polarity sign.)
(2) Change of function and so on
Executes change of various kinds of functions in Function mode, and also changes of function
during Check mode.
●
During the operation in the paragraph 6
−
1
−
2, the key is pressed for
more than 2 s, each key operation will be executed continuously at a
constant interval without the operation of ON/OFF.
6
−
2. “A/Z OFF
Y
” key
6
−
2
−
1. Operation in Measurement mode
Only when the Tare weight cancellation
(
A/Z
)
function is operating, tare weight cancellation
clear
(
A/Z OFF
)
operates and at the same time when load display value becomes gross weight
display value, “A/Z” on status display will put off.