69
Procedures
・
Check the input signal for external control
Press the “ENTER” key.
The load display will show “IN”.
By pressing the “ENTER” key again, the load
display will change “
−
IN
−
0”.
At this time, ON/OFF condition for external
control input signal can be monitored on LED
for status display, and on load display section.
5
RUN display
:
ZERO input
A/Z display
:
A/Z input
CHECK display
:
A/Z OFF input
HOLD display
:
HOLD input
LOCK display
:
LOCK input
Load display 10
0
digit
:
SEL.1 input SEL.2 input
Load display 10
0
digit is “0”
:
Open both SEL.1 and SEL.2.
Load display 10
0
digit is 1.
:
Short only SEL.1
Load display 10
0
digit is 2.
:
Short only SEL.2
Load display 10
0
digit is 3.
:
Short both SEL.1 and SEL.2.