51
Item
Function No. Setting value
Contents
Operation condition for
accumulating function
F-50
000 ~ 121
●
000
Accumulation command operation.
10
0
digit: 0 = only in stale, 1 = Always
Automatic accumulating operation
10
1
digit: 0 = OFF,
1 = Synchronizing with finish,
2 = Automatic accumulation
at
stable
Automatic accumulating condition
10
3
digit: 0 = Only in OK mode,
1 = No condition
Target of brand changeover
F-51
●
0
Changeover by function
1
Changeover by external control input
2 Changeover
by
communication
Brand number
F-52
●
0
Brand 0
1 Brand
1
2 Brand
2
3 Brand
3
4 Brand
4
5 Brand
5
6 Brand
6
7 Brand
7
Setting F key operation
F-55
●
0
OFF
1 Print
Input
2 Hold
Input
3 Batch
start
4 Emergency
stop
5 Zero
clear
6 Tare
clear
7 Accumulation
clear
8 Measurement
comparison
9 Forced
batch
finish
External control input INPUT1
operation setting
(Settings 1 to 8 operate all the
time,
settings 9 to 27 operate only in
measurement mode.)
F-60
●
0
OFF
1
key operation
2
key operation
3
key operation
4
key operation
5
key operation
6
key operation
7
key operation
8
key operation
9 Zero
clear
10 Tare
clear
11 Flow
start
12 Discharge
start
13
Flow/Discharge changeover
ON: Discharge, OFF: Flow
14
Clear the last accumulated data
15 Accumulation
clear
16 Error
cancellation
17 Print
command
18 Hold
●
Has been set as default
Summary of Contents for CSD-903-EX
Page 1: ...DIGITAL INDICATOR CSD 903 EX Instruction Manual EN294 1492 H ...
Page 2: ...IV ...
Page 91: ...75 ...
Page 93: ...77 ...
Page 95: ...79 ...
Page 97: ...81 ...
Page 199: ...183 Inform us about the contents of failure and situation at site in details ...
Page 204: ...188 Inform us about the contents of failure and situation at site in details ...
Page 206: ...190 Inform us about the contents of failure and situation at site in details ...
Page 209: ...193 Inform us about the contents of failure and situation at site in details ...
Page 224: ... ...
Page 225: ......