Connecting method
3-1.
Connector pin configuration for communication
Refer to 2-1. Rear panel for CC-Link I/F (1), Connector for communication.
3-2.
Notes of Connection
z
When the wiring, be sure to the instrument power supply is OFF.
z
Do not supply the AC power until complete the installation. This instrument does not have power
switch (ON/OFF).
z
Do not fell or make a strong impact on this instrument rear panel terminal block because it is made
of resin.
z
Striped electrical cable tip length is 6 mm.
z
The tightening torque of terminal screws on the terminal block is 0.5
~
0.6 N
・
m.
z
Cables which connecting this instrument isolate from noise sources, for example, power supply line
and I/O for control's as much as possible.
z
Be sure to connect the ground wire must be D single ground. Do not common the ground with a
kind of power supply.
z
For CC-Link cable connection, use twist pear cable wire with shield (Cable for CC-link) and connect
the shield in terminal block's SLD terminal or F.G.terminal.
•
Connect the termination to the CC-Link connector to electrical termination which is far from PLC as possible.
•
Use the connecting cable for CC-Link.
•
Refer to the “Construction and specification of network system” from the latest version of CC-Link Cable Wiring Manual
published by CC-Link Partner Association about communication speed and cable length