Medical Fridge

CONDENSER CLEANING

Always follow these steps when cleaing the unit condenser.

VOLTAGE STABILIZER 

The Refrigerator is manufactured with a built in voltage stabilizer 
that protects the unit from damage due to voltage fluctuations

 Step 1: Unplug

Unplug the unit from the mains 
electricity supply

 Step 2: Remove Grille

Remove the bottom grille 
underneath the door(s) by 
lifting slightly and pull away 
from the retaining handle. No 
tools required. 
CAUTION: Some pipework & 
components might be hot

 Step 3: Condenser

Locate the condensor next to, 
and attached to, the fan.
*It looks similar to a car 
radiator

 Step 4: Soft Brush

Take a soft brush and clean 
the fins. 
CAUTION: The condensor is 
very delicate

• 

Do not to bend the fins.

• 

Never poke anything 
through the fins

• 

Never use water

• 

Never use a hard brush

 Step 5: Vaccuum

If possible, use a vaccuum to 
clean away the debris.
Check the rest of the 
compartment for any foreign 
objects and remove.

 Step 6: Remove Tools

Ensure that all tools have been 
removed.

• 

Never use a wire or hard 
brush

 Step 7: Power

Plug the unit back in to the 
mains electricity supply.

 Step 8: Data Log

Log the date of the cleaning 
and make a dairy note to clean 
it again in 2 to 3 months

• 

Never use a wire or hard 
brush