E-SOUM-PLUS Rev E
4
Electromagnetic Compatibility Guidance (in accordance with UL/EN/IEC 60601-1)
Guidance And Manufacturer’s Declaration – Electromagnetic Emissions
(Table 201)
The SonicOne CLINIC SYSTEM is intended for use in the electromagnetic environment specified below. The customer or the
user of SonicOne CLINIC SYSTEM should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The SonicOne CLINIC SYSTEM uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The SonicOne CLINIC SYSTEM is suitable for use in all
establishments other than domestic and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Table 1.1 Guidance & manufacturer’s declaration on electromagnetic emissions (EN table 201)
NOTE: The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals
(CISPR 11 class A).
NOTE: The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals. If it is
used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer
adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such
as relocating or re-orienting the equipment.