CONNECTIONS
PRECAUTIONS
FEA
TURES
PREP
ARA
TION
TROUBLE-
SHOOTING
OTHERS
INSTRUCTIONS FOR MEDICAL USE
<according to the Medical Safety/EMC standard EN 60601-1-2>
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to
be installed and put into service according to the EMC information provided in the ACCOMPANY-
ING DOCUMENTS.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
Electro Static Discharge (ESD)
1 Electro Static Discharge (ESD) warning symbol
2 The areas identified with this symbol should not be touched as much as possible because they
are sensitive to ESD and may cause equipment malfunction and failure due to ESD. When it
is inevitable to touch these areas in installation, maintenance, or inspection, the following ESD
precautionary procedure should be used.
3 ESD precautionary procedure
(1) The power plug should be disconnected for safety’s sake first.
(2) The worker should wear an antistatic strap and attach it to the ground pin of the power plug
or the metal sheet on the bottom surface of the product. This eliminates the potential difference
between the product and the worker to prevent electro static discharge. (When no antistatic
strap is available, the worker may work while touching the ground pin of the power plug or the
metal sheet on the bottom surface of the product with one hand.)
(3) The necessary work for installation, maintenance, or inspection should be performed under
the above-mentioned conditions.
4 It is recommended that all staff involved in the work in which they may touch the areas identified
with the ESD warning symbol should receive an explanation of the ESD warning symbol and
training on the ESD precautionary procedures.
5 Required minimum contents in the explanation and training on the ESD precautionary
procedures
(1) Explanation on the principle of occurrence of static electricity and the ESD precautionary
procedures should be provided.
(2) Training on the ESD precautionary procedures using the actual product should be
provided.
Technical description
List of all cables and maximum length of the cable and other ACCESSORIES
AC power cord
USB cable
PAPER/INK
RIBBON SET
Maximum length
2 m
2 m
Reference page in this operation manual
This page, the previous pages for safety and page
22 for accessories
This page, the previous pages for safety
Page 12 and 23 for PAPER/INK RIBBON SET
WARNING:
The use of ACCESSORIES and cables other than those specified, with the exception of cables
sold by the manufacturer of the Model CP30DW as replacement parts for internal components,
may result in increased EMISSIONS or decreased IMMUNITY of the Model CP30DW.
WARNING:
The Model CP30DW should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use is necessary, the Model CP30DW should be observed to verify normal
operation in the configuration in which it will be used.
Summary of Contents for CP-30DW
Page 32: ......