10
For the functions of the MODE switch, refer to "MODE SWITCH
FUNCTIONS" on page 21 - 23.
Medical Video Signal
1
Turn off the power switches of the Video copy processor and
the equipment to be connected.
2
Connect the VIDEO input terminal of the Video copy
processor to the video output terminal of the connecting
equipment.
Rear panel
Medical
equipment
AC LINE
1 2 3 4 5 6 7 8 9 10
OFF←
ADJUST
H-POSI AFC
REMOTE
IMP TRAP GAIN IMAGE AFC DIR MEMORY SCAN SAVING PAPER
75
Ω
ON ON NEGA ON REV FIELD OVER ON SUPER
HIGH OFF OFF POSI OFF NOR FRAME UNDER OFF HD
1 2 3 4 5 6 7 8 9 10
ON
OFF
VIDEO
IN
OUT
Set the MODE switch as follows:
(This is standard setting.)
SW-NO.
Setting
75
Ω
OFF
OFF
POSI
OFF
NOR
FRAME
UNDER
OFF
HD
1
2
3
4
5
6
7
8
9
10
1 2 3 4 5 6 7 8 9 10
OFF←
REMOTE
IMP TRAP GAIN IMAGE AFC DIR MEMORY SCAN SAVING PAPER
75
Ω
ON ON NEGA ON REV FIELD OVER ON SUPER
HIGH OFF OFF POSI OFF NOR FRAME UNDER OFF HD
1 2 3 4 5 6 7 8 9 10
ON
OFF
※
Setting of Switches
2
Select
γ
-curve " " and " " - " " for ultrasonic diagnosis equipment.
Especially " " is recommended. Refer to the
γ
-curve set mode on page 15.
2
※
Refer to the " MODE SWITCH FUNCTIONS " on page 21 .
To VIDEO
output
terminal
To VIDEO
IN terminal
Connect
Power
cord