2 - 5
2 SETUP
No. 99MBC034B
(3) If the EV-16Z is used, push in the spindle of the connected
gages to pass through the origin one after another, and each
time it passes through the origin the corresponding bar dis-
play of the CH starts flashing.
[Display on the D-EV]
+NG
UNIT
GO
-NG
(4) If the EV-16Z is used, push in the spindles of all connected
gages to pass through the origins, then the counter enters the
counting mode.
[Display on the D-EV]
UNIT
+NG
GO
-NG
c) Precautions in using the EV-16Z
If the EV-16Z is used, note the following precautions.
IMPORTANT Be sure to pass the spindle through the origin for
detection properly. If the spindle vibrates near the
origin, the origin may not be detected correctly.
TIP
• The origin-embedded gage has a specific origin in-
side the gage. When the spindle passes through the
origin the signal generates, then the preset position is
restored.
• Origin detection is usually performed when the power
is turned on. If the origin re-detection parameter
(P.No.42) is set to 1, the counter enters the origin de-
tection wait mode after completing preset or tolerance
setup (for details, refer to the section, “3.1 Setting
Parameters (with D-EV)”), or when the HOLD signal
is triggered.
• If the HOLD signal is input again during origin re-de-
tection, the counter resets origin re-detection, except
when resetting an error.