Hana
®
Orthopedic Surgery Table Owner’s Manual
Mizuho OSI 2017
137
NW0508 Rev N
The pads are intended to be cleaned in place. They do not need to be rotated or removed.
Clean with standard hospital disinfectants labeled for use on table pads. Always dilute and rinse per
manufacturer’s label instructions. Wipe dry with a lint-free cloth. DO NOT soak or autoclave the pads.
Clean the pad on the tabletop, wipe dry with a lint free cloth, and then reposition the pad to lie flat on the
tabletop.
NOTE: The use of bleach or highly concentrated chemicals may discolor the cover and will void
the warranty on the pad.
When cleaning the bottom of the pad on the tabletop, simply lift one end of the pad, and fold it over onto
the other end. Clean the pad on the tabletop, wipe dry with a lint-free cloth, and then reposition the pad to
lie flat on the tabletop.
10.1.3 Femoral Hook Support and Femoral Hook
Prior to each patient use, inspect, clean, and sterilize the Femoral Hook(s) and Femoral Hook Support
according to the hospital’s protocol for reprocessing surgical instruments.
Do not use steel wool, wire brushes, pipe cleaners or abrasive detergents. Use of anything other than high
quality brushes designed for surgical instrument cleaning may result in damage.
Disassembly of the Femoral Hook Support before cleaning is not necessary unless severely soiled or
dictated by hospital policy.
Do not use high acid (pH 4 or lower) or high alkaline (pH 10 or higher) disinfectant products, such as
bleach and bi-chloride of mercury.
10.1.4 Hana
®
Knee Flexion System
®
Prior to each patient use, inspect, clean, and sterilize the Knee Flexion System
®
according to the hospital’s
protocol for reprocessing surgical instruments.
Do not use steel wool, wire brushes, pipe cleaners, or abrasive detergents. Use of anything other than high
quality brushes designed for surgical instrument cleaning may result in damage.
Disassembly of the Knee Flexion System
®
before cleaning is not necessary unless severely soiled or
dictated by hospital policy.
Do not use high acid (pH 4 or lower) or high alkaline (pH 10 or higher) disinfectant products, such as
bleach or bi-chloride of mercury.
10.2
Sterilization
The Femoral Hook Support, Femoral Hook(s), and Hana
®
Knee Flexion System
®
must be sterilized before
use. Sterilization of these components is accomplished by steam. To achieve a sterility assurance level of
10
-6
, Mizuho OSI recommends the sterilization parameters listed below: