KBS – Read Keyboard
Lock/Unlock Status
Use the keyboard status (KBS) command to determine whether the CAL (or
ATM), Gauge, Degas, and Trip Points buttons are locked or unlocked.
KBS command from host:
#01KBS
↵
Response from module:
*01 1 KB OFF
↵
• The “1 KB OFF” response means the ATM (or CAL), Gauge, Degas, and
Trip Points buttons are unlocked.
• The “1 KB ON” response means the ATM (or CAL), Gauge, Degas, and
Trip Points buttons are locked.
Table 3-16 Module Status RS-485 Strings
Response
Cause
Solution
00 ST OK
Module is operating normally.
No solution necessary.
01 CGBAD
Conductron sensor is inoperable.
• Cycle power to the module.
• If necessary, replace the gauge assembly (see
05 IG HV
Micro-Ion gauge grid voltage failure.
Cycle power to the module or send IG1 command
to clear the status response (see page 61).
06 IG EM
Micro-Ion gauge emission failure.
07 IGFIL
One Micro-Ion gauge filament is open.
• If SF command is set to AUTO, ALT, or Y,
operation automatically switches to the other
filament (see page 52).
• If SF command is set to MAN or T, send IG0
command, then send IG1 command to switch
filaments (see page 50).
08 POWER
Power cycle has occurred, module is starting up.
No solution necessary. The “
00 ST OK
” string will
be generated next if the module is operating
properly.