5.
Basic Set-Up and Operational Parameters
5.1.
Basic Set-Up
The basic settings described below are in general non-recurring and must only be done upon
the very first commissioning of the unit. The basic menu includes the selection of the desired
operation mode and the desired brightness of the LED display.
To start the basic set-up procedure, keep down both front keys at a time for at least 3 seconds.
Menu
Selection
Text
Description
Mode
Operating mode of the unit
Single
Single channel (input A only)
Dual
Dual channel (input A and input B separately)
A u B
Summing mode (input A + input B)
A – B
Differential mode (input A – input B)
A d B
Ratio mode (input A : input B)
A m B
Multiply mode (input A x input B)
Bright
Brightness of display ( 20, 40, 60, 80 or 100 % )
Update Update time
Updates the display every x.xxx seconds.
Code
Access code for the programming keys
no
Keys are open for programming
ALL
Keys are locked for all operations
Linear Linearisation
mode
no
Linearisation is off, all linearisation parameters
are irrelevant
1-qua
Linearisation in the range 0 – 99999.
4-qua
Linearisation in the range –99999 bis +99999.
A-CHAr Analogue output characteristics
-10_10
Full voltage range, -10V bis +10V
0 _ 10 Positive voltage range only, 0 – 10V
0 _ 20 Current range 0 – 20 mA
4 _ 20 Current range 4 – 20 mA