Ax34601a_e_A5.doc / Feb-06
Page 14 / 20
The result of the calculation <AB> can then have a final scaling to receive user-friendly
engineering units, by means of the following scaling factors:
Menu
Selection
Text
Description
m_Fac
Proportional
Factor
Multiplies the result by this setting
d_Fac Reciprocal
Factor
Divides the result by this setting.
P_Fac Additive
Factor
Adds this setting to the result, under consideration
of the sign
dPoint Decimal
Point
Sets the decimal point to the final display
Final Display Value
=
Result from combination <AB>
x
m_Fac
d_Fac
+/-
P_Fac
5.2.4.
Parameters for scaling of the analogue output
The analogue output depends only from the value shown in the display of the unit, and from the
scaling parameters shown below. Therefore the analogue signal would also contain possible
combinations and calculations, and potential linearisation as well.
Total output swing and zero output have already been defined by the basic setup parameters.
The operating range or window of the output can be set by means of the following parameters:
Menu
Selection
Text
Description
AnAbeG
AnAEnd
Analogue beginning
Analogue end
These parameters will map a user-defined window
of the total display range onto the selected
analogue output range.
Example: Anabeg = „-1500“ and Anaend = „2100“
will generate the analogue zero output signal at a
display of „-1500“ (see Offset), and the analogue
full scale output signal at a display of „2100“
(see Gain)