External configuration
In order for the safety switch functionality to work, some parameters of the
wheelchair electronics need to be set correctly:
•
Curtiss-Wright R-net parameters on page 16
•
Dynamic Controls parameters on page 17
Curtiss-Wright R-net parameters
For more information, please refer to the
Curtiss-Wright manual SK77981-14
| R-net Technical Manual | Chapter 7 - Controls
.
External Profile Jack Detect
Controls
>
Global
>
External Profile Jack Detect
VALUE
DESCRIPTION
Off
The Twister Pro will function as a normally open switch.
On
The Twister Pro will be able to detect both open-
and short-circuits in the wiring. If such conditions are
detected, the function associated with the
External Pro-
file Switch Jack
will be disabled and a warning will be
displayed on the screen.