Programming
EPM2000/EPM2000e Reference Guide
42
RS-232 B
AUD
R
ATE
Set or query the RS-232 baud rate of the EPM2000.
Command Form
COMMunication RS232 BAUD <?|1200|2400|4800| 9600|19200|38400>
Examples
Input
Instrument Reply
comm rs232 baud 2400
none
comm rs232 baud ?
COMM: RS232: BAUD 2400
RS-232 P
ARITY
Set or query the RS-232 parity bit setting of the EPM2000.
Command Form
COMMunication RS232 PARity <?|ODD|EVEN|NONE>
Examples
Input
Instrument Reply
comm rs232 par even
none
comm rs232 par ?
COMM: RS232: PARITY EVEN
RS-232 S
TOP
B
ITS
Set or query the number of stop bits the EPM2000 injects into the communications stream.
Command Form
COMMunication RS232 STOPbits <?|1|2>
Examples
Input
Instrument Reply
comm rs232 stop 2
none
comm rs232 stop ?
COMM: RS232: STOPBITS 2