Programming
EPM2000/EPM2000e Reference Guide
47
W
AVELENGTH
C
OMPENSATION
Set or query the Wavelength Compensation value. Argument is in meters in scientific notation.
Command Form
CHannel <A|B|ALL> WAVElength <?|ON|OFF>
Examples
Input
Instrument Reply
ch a wave 5.47e-7
none
ch b wave 5.14e-7
none
ch all wave ?
CHAN A: WAVELENGTH
COMPENSATION: 547 nm
CHAN B: WAVELENGTH
COMPENSATION: 514 nm
Z
ERO
In thermopile mode, this command sets the specified channel’s current probe input voltage as a zero
baseline value. While taking a statistics measurement, the EPM2000 channel will reset the batch
upon receiving this command.
Command Form
CHan <A|B|ALL> ZERO
Examples
Input
Instrument Reply
ch a zero
<
batch resets/
thermopile zeroes>
Summary of Contents for EPM2000
Page 2: ...EPM2000 EPM2000e Dual Channel Joulemeter Power Meter...
Page 3: ......
Page 4: ...EPM2000 EPM2000e Dual Channel Joulemeter Power Meter Reference Guide...
Page 23: ...Features Description EPM2000 EPM2000e Reference Guide 14...
Page 39: ...Operation EPM2000 EPM2000e Reference Guide 30...
Page 73: ...Specifications EPM2000 EPM2000e ReferenceGuide 64...
Page 75: ...Appendix Declaration of Conformity EPM2000 EPM2000e Reference Guide 66...