Touch Screen
Figure 1-1:
AquaMax
®
touch screen
Instrument operation is controlled through a touch-screen user interface.
The touch screen displays instrument status, provides access to
instrument settings and programs, and controls the optional StakMax
Plate Stacker.
Wash Heads
AquaMax wash heads have either 96 or 384 paired probes. The longer
probe of each pair is used for aspiration. The short probe is used for
dispensing. The plate wash heads have straight dispense probes and are
intended for thorough cleaning and aspiration. The cell wash heads have
angled dispense probes to minimize cell loss during washing.
The wash heads are interchangeable and can be removed and
re-attached using a simple lever mechanism.