Command Format
Document Conventions
All commands must follow the format:
/
CMD_NAME
[
command_args
]
CR
•
‘/’ (Forward slash) indicates start of a new command and clears
firmware command buffer. It is optional and can be omitted.
•
CMD_NAME
is a command name separated from command_args
by a white space.
•
Command_args
is a list of zero or more arguments, each
separated by white space.
•
CR
is the end-of-command character
Commands and arguments are not case sensitive.
It is acceptable, but not required, to allow commands to be abbreviated
with a minimum 4 characters to uniquely identify the command. The
commands will be searched in alphabetical order. The first match will be
the selected command.
String arguments must not include the prohibited characters: CR LF TAB /
< > \ ? * : | “ .
String arguments can be enclosed in double quotes.
Table E-1:
Command Formatting
CAPS ITALICIZED
Command
names
|
Indicates the possible choices for that argument
,
A comma separates the arguments. In the instrument,
arguments are space delimited; therefore the comma is
shown here only for clarity
italicized
Command arguments
ON
An argument in all capital letters, such as ON, implies that a
literal string is required.
CRLF
Carriage Return and Line feed characters (‘\r\n’)
Summary of Contents for AquaMax 2000
Page 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Page 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Page 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Page 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Page 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Page 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Page 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...