AquaMax
®
2000/4000 Microplate Washer User Guide
50
0112-0151 K
Wash Settings
The Wash step washes a microplate using predefined parameters.
5
Probe height–final probe height above the bottom of the well. The
range is 0 to the top of well. The maximum probe height is the well
depth specified in the plate library.
To avoid damage to probes and plate bottom, the AquaMax
®
Microplate Washer has a “no touch” feature. If the height is set to
0mm, the aspiration pins are offset from the well bottom by a
minimum clearance.
Table 4-11:
Aspiration Rates
AQ 4000
AQ 2000
96 (
μ
L/well/sec) 384 (
μ
L/well/sec)
5
Fast
270
70
4
240
60
3
200
50
2
Gentle
80
20
1
50
10
Table 4-10:
Aspirate settings
Item
Description
Tip:
For plate washing, minimal residual volume is best achieved using
Crosswise aspiration, a rate of 5, a fast descent speed, a 1 second dwell time,
and 0 mm to 0.2 mm probe height.
Note:
To prevent overflow when using the Wash function with a plate wash
head, add a surfactant such as Tween to wash solutions.
Summary of Contents for AquaMax 2000
Page 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Page 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Page 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Page 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Page 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Page 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Page 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...