Digidata 1550 Low-Noise Data Acquisition System User Guide
32
5023399 A
AC Power
The AC Power is supplied from a single AC input connector on the rear
panel for the supplied AC power line cord.
Operational and Environmental
Table A-7:
AC Power
Voltage
Description
Voltage input rating for 1550
100 V–240 V AC, 50/60 Hz
50 W Max.
Replaceable fuse type
Cartridge, fast acting
Fuse dimensions
6.3 mm x 35 mm
Fuse rating
1.0 Ampere, 250 V
Table A-8:
Operational and Environmental Specifications
Specification
Measurement
Weight
3.6 kg
Dimensions (H x W x D)
109 mm x 483 mm x 363 mm
Mains Power Input
100 – 240 VAC, 50/60 Hz, Max. 50 W
Mains Voltage Fluctuations
Not to exceed 10% of nominal supply voltage
Equipment Class
1
Pollution Degree
2
Installation Category
2
Operating Environment
Indoor Use Only
Altitude
Not to exceed 2000m
Operating temperature:
15 to 30 °C (59 to 86 °F)
Humidity
<80% RH at 25 °C (77 °F)
Ingress Protection:
IP20