EMax® Plus Microplate Reader

Quick Start Guide

The EMax® Plus Microplate Reader is a filter-based absorbance microplate reader, with
eight standard filters for a wide range of applications. An external computer running the
SoftMax® Pro Microplate Data Acquisition and Analysis Software provides integrated
instrument control, data display, and statistical data analysis.

Setting Up the Instrument

Before setting up the EMax® Plus Microplate Reader, prepare a dry, flat work area that
has sufficient space for the instrument, host computer, and required cables. To provide
access for disconnecting power from the instrument, maintain a 20 cm to 30 cm (7.9 in.
to 11.8 in.) gap between the rear of the instrument and the wall. To ensure sufficient
ventilation, do not block the ventilation grid on the right side of the instrument.

Note:

Make sure that the host computer meets or exceeds the system

requirements for the SoftMax Pro Software. See the SoftMax Pro Software
installation guide.

Setting up the instrument includes the following procedures:

Installing the SoftMax Pro Software. See the SoftMax Pro Software installation
guide.

Removing the Shipping Lock From the Microplate Carrier, see page 2
Connecting the Instrument Cables, see page 2
Starting the Instrument and the Software, see page 3
Selecting the Instrument in the Software, see page 4

Contents of the Package

The package contains the instrument and the tools and accessories that are required for
setting up the instrument as follows:

EMax Plus Microplate Reader
EMax Plus Microplate Reader Quick Start Guide
SoftMax Pro Software Product Key
USB computer connection cable
AC power adapter

For a complete list of the contents of the package, see the enclosed packing list.

Downloading the SoftMax Pro Software

Molecular Devices recommends that you install the SoftMax Pro Software on your host
computer before setting up the instrument.

To download the latest version of the software, visit:

www.MolecularDevices.com/SMPdownloadsite

Summary of Contents for EMax Plus

Page 1: ...equirements for the SoftMax Pro Software See the SoftMax Pro Software installation guide Setting up the instrument includes the following procedures Installing the SoftMax Pro Software See the SoftMax Pro Software installation guide Removing the Shipping Lock From the Microplate Carrier see page 2 Connecting the Instrument Cables see page 2 Starting the Instrument and the Software see page 3 Selec...

Page 2: ...1 Make sure that the instrument and host computer are placed on a dry flat work area with sufficient space for both devices and the required cables To ensure sufficient ventilation and provide access for disconnecting power from the instrument maintain a 20 cm to 30 cm 7 9 in to 11 8 in gap between the rear of the instrument and the wall 2 Make sure that the power switch on the rear of the instrum...

Page 3: ... SoftMax Pro Software installation guide Molecular Devices recommends turning the power off when the instrument is not in use Activating Your Software License When you first start the SoftMax Pro Software the Software License Activation dialog is displayed If you are running a trial version of the software click the Help tab in the ribbon and then click Software License To activate your SoftMax Pr...

Page 4: ...plication help or user guide The instrument icon shows the connection status of the instrument Icon Status Instrument Connected Instrument Disconnected Instrument Simulated Table Q 1 Instrument Connection Status If the instrument status shows as disconnected check to make sure that the instrument is powered on and that the connections between the instrument and the computer are secure To select th...

Page 5: ...s on page 5 Setting Up an Absorbance Endpoint Experiment on page 7 Loading and Unloading Microplates The microplate carrier on the EMax Plus Microplate Reader moves in and out of the read chamber during the read and then returns to the home position after the read ends If the microplate carrier does not return to the home position you can initialize the carrier to return it to its home position by...

Page 6: ...rument can read 96 well clear bottom microplates and strip wells For most applications use microplates with flat bottom or round bottom wells Molecular Devices recommends that you do not use V bottom microplates unless an application specifically requires their use Irregular plastic density in the tip of the well can cause inaccurate measurements The type of microplate and the way it is handled ca...

Page 7: ...lect a contiguous rectangular region on the microplate drag the cursor to select the wells to be read Columns do not need to start with 1 but must be contiguous To read all the wells in the microplate select the Select All check box Note If you select a subset of wells the instrument reads the entire microplate but reports the data for the selected wells only 8 Click OK 9 Save the settings as a da...

Page 8: ...resentatives You can contact your local representative or contact Molecular Devices Technical Support by telephone at 800 635 5577 U S only or 1 408 747 1700 In Europe call 44 0 118 944 8000 Please have the system ID number system serial number software version number and the system owner s name available when you call For research use only Not for use in diagnostic procedures The trademarks menti...

Reviews: