FlexStation 3 Benchtop Multi-Mode Microplate Reader User Guide
156
0112-0127 B
Table B-3:
Absorbance Photometric Performance
Item
Specification
Wavelength range
200 nm to 1000 nm
Wavelength selection
Monochromator tunable in 1 nm increments
Wavelength bandwidth
4.0 nm
Wavelength accuracy
± 3.0 nm
Wavelength repeatability ± 0.2 nm
Photometric range
0.0 OD to 4.0 OD
Photometric resolution
0.001 OD
Photometric
accuracy/linearity,
0 OD to 2.0 OD
± 0.006 OD ± 1.0%
Photometric precision
(repeatability),
0 OD to 2.0 OD
± 0.003 OD ± 1.0%
Stray light
0.05% at 230 nm
Photometric stabilization Instantaneous
Photometric drift
None
Continuous referencing of monochromatic input
Calibration
Automatic before first kinetic read and before every
endpoint read
Optical alignment
None required
Light source
Xenon flash lamp (50 Watts)
Average lamp lifetime
1 billion flashes
Photodetectors
Silicon photodiode
FlexStation_manual.book Page 156 Tuesday, June 25, 2013 10:22 AM