Performance Specifications
0112-0127 B
159
Table B-10:
Temperature Regulation
Item
Specification
Reading chamber
Isothermal when temperature regulation is not
enabled
Range
2°C above ambient to 45°C when temperature
regulation is enabled
Ambient temperature must be >20°C to achieve
temperature regulation at 45°C
Resolution
± 0.1°C
Accuracy
± 1.0°C for reading chamber
Temperature uniformity
at equilibrium
± 1.0°C at 37°C
Chamber warm-up time
30 minutes (measured on air) after initiation of
temperature regulation
Temperature regulation
4 sensors
Temperature regulation
diagnostics
Temperature regulation system is continuously
monitored and updated
Evaporation
Plate lid required to minimize evaporative cooling
Recommended
microplate
Flat-bottom microplates with isolated wells and lid
Control
Reports reading chamber temperature only
Table B-11:
Automix with SoftMax Pro Software
Item
Specification
Plate mixing modes
Selectable: off, once before any reading, or once
before and between kinetic readings
Plate mixing duration
Selectable: 0 to 999 seconds (three-second default)