FlexStation 3 Benchtop Multi-Mode Microplate Reader User Guide
136
0112-0127 B
Assessing a Mechanical Problem
Depending on when the problem occurs during the fluidics operation,
one of several error messages can appear in the SoftMax Pro Software.
The error message might not provide complete information about what
has occurred. The software cannot confirm the presence, absence, or
exact position of tips or plates. Whenever you experience a mechanical
failure of the fluidics operation, follow the entire procedure until you can
pinpoint the source of the problem.
•
Fluidics module not detected.
This message might occur if the fluidics module is not correctly
installed.
•
Pipettor failure. There may be undispensed fluid in the pipette
tips.
This message might occur if the tip rack is jammed. Tips might be
missing. To help prevent frequent tip dropping, periodically clean
the pipettor barrels. See
Cleaning the Barrels on the Pipettor
•
Pipettor failure. Undispensed fluid in the tips.
This message occurs if fluid remains in the tips. The instrument
needs to be manually recovered at this point.
Note:
If you need to contact technical support during the following
procedures, be sure to note which error message you observed.