FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
103
Export Statistics
Use this tab to configure kinetic reductions,
for example
,
averages and
maximum and minimum values, for selected numbers of reads within
each well. Multiple kinetic reductions can be configured, each resulting
in another export file. Within each file only one value is reported per
well.
The table below describes the settings and options of this dialog box:
Item
Description
Enable Statistics
Check this box to have a statistic file created.
File Name Options
Choose a name for the exported statistic file.
Use name of data file
—Export files with the data or
protocol file name, with *.stat
n
extension, where
n
is an
integer
1.
Use user-defined name
—Enter your own name for the
export files (maximum of 25 characters). Files are given
*.stat
n
extension, where
n
is an integer
1.
Data Format
Format the output from this group of options:
Show Labels—
When checked, the output file contains
information about the processing options selected (for
example
,
name of *.fmd file and well labels). Any
corrections done prior to exporting the data (
e.g.,
negative control correction) are indicated in text format
at the top of the exported ASCII file.
Order of Well Data—
Order the data by column (for
example, A1, A2, A3) or by row (
e.g.,
A1, B1, C1). Set
this option in accordance to the way your spreadsheet or
database handles well data.
Misc. Parameters
Select experiment parameters and group statistics to
export with the data.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...