ScreenWorks Software Overview
118
0112-0109 H
The table below describes the options available on the
Mix Fluid
process page:
Item
Description
Mix Plate
Select the plate that has the fluid to be mixed. Typically,
it is a source plate from which you will be transferring
fluid.
Quadrant
If appropriate, select a quadrant from which fluid is
mixed.
Note:
Quadrant selection is only available when
liquid is dispensed to a plate using a pipettor
head of lower density, for example, 384-well plate
using a 96 pipettor head.
Volume
Enter volume (in
μ
L) to mix.
Speed
Enter speed (in
μ
L/s) to mix the fluid.
Height
Enter distance (in
μ
L) from the bottom of the well that
the tips will be inserted prior to mixing.
Note:
Tips move up and down with aspirate and
dispense commands. These movements help to
prevent well overflow when the tips are
submerged.
Expel Volume
Enter an additional volume (in
μ
L) the pipettor
dispenses, beyond the volume entered in the
Volume
field above. This helps to ensure that all fluid is expelled
from the tips after mixing.
Strokes
Enter number of times to aspirate and dispense mix
volume. One stroke is equal to one aspiration and one
dispense.
Removal Speed Enter rate (in mm/s) at which tips are pulled from well
immediately after the mix. This value is adjusted for
fluid viscosity or fluid volume in the well.
Pause In Well
Enter time (in seconds) to pause with the tips in the well
after mixing and before the next pipettor head move.
Unload Tips
After Mix
Automatically unloads tips after the mix step.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...