Wash Pins (Pin Tool)
The table below describes the options available in the
Wash Pins
process page, for a
Pin Tool
:
Settings/Options
Description
Fluid Type
Select the fluid in the wash reservoir during tip
washing, from wash fluid container A or B.
Pump Speed
Select the speed (slow or fast) you want the pump
to fill and empty the wash reservoir.
Note:
Molecular Devices recommends you
use
Fast
except with volatile solvents.
Wash Cycles
From the list, select the number of times that fluid
cycles through the wash reservoir (up to 5). For
each cycle, the pipettor executes the number in the
Strokes
field.
Strokes
From the list, select the number of times to raise
and lower the pins. One stroke is equal to one up-
stroke and one down-stroke.
Down Speed
Set the speed of the pin tool during the down-
stroke.
Stroke Distance
Set the height of one stroke that the pin tool
travels.
Up Speed
Set the speed of the pin tool during the up-stroke.
Removal Speed
Set the speed that the pin tool is withdrawn from
the wash location. The faster the withdrawal speed
the more fluid remains on the pins.
Pause in Well
Set a time to hold the pin tool at the bottom of the
wells in the wash solution after the wash is
completed.
Unload Pin Tool After
Wash
Check to have the pin tool removed after washing
is complete.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...