0112-0109 H
149
7
Calibration and Signal Test
Chapter 7 provides the procedures and information required to
calibrate the optics and run a signal test of the FLIPR
®
Tetra System.
Optical Calibration
Calibrating the optics of FLIPR
®
Tetra System requires taking an image
of the 96-, 384-or 1536-well plate and using it as a mask to locate the
wells. Typically, calibrations are performed with the following
procedures:
After LED banks or pipettor head has been exchanged.
After a new plate has been added to the plate library.
If no filters are in the instrument, install them at this time (see
Adding a Read Plate to Plate Library
The FLIPR
®
Tetra System has default plate masks for 96-, 384-, and
1536-well read plate types. If you want to use a plate type that varies
significantly from the default formats, you must create new dimensions
and mask for that plate type. Use the description of the Plate Library
menu item, (see
Plate Process Definition on page 62
), as a reference
guide for options and parameters available when adding a plate to the
software.
To add a plate and mask to the software:
1.
Prepare a 1 x 10
-8
M fluorescein solution.
2.
Ensure the microplate bottom is clean and free of scratches and
dust.
3.
Dispense enough fluorescein solution into each well so as to
cover the bottom. Dispensing half of the total well volume is
typically a safe volume with which to work. This may be done
using FLIPR
®
Tetra System pipettor.
4.
Place the microplate containing fluorescein in the read position
and select
Plate Library
from the
Tools
menu.
5.
From the
Plate Process Definition
list, select a plate with a
similar well format.
6.
Click
Copy Plate
.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...