FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
211
214: “The 2 upper LED
banks are not identical
wavelengths. Please
install identical
wavelengths to both
upper positions.”
The upper pair of LED
banks has mismatched
LED module
wavelengths.
Make sure LEDs of the
same wavelength are
located in the upper LED
bank position on either
side of the read plate.
215: “The 2 lower LED
banks are not identical
wavelengths. Please
install identical
wavelengths to both
lower positions.”
The lower pair of LED
banks has mismatched
LED module
wavelengths.
Make sure LEDs of the
same wavelength are
located in the lower LED
bank position on either
side of the read plate.
217: “One or more LED
banks not detected.
Please verify all four
banks firmly installed.”
Cannot detect one of
the LED modules.
Make sure LED module is
properly connected.
Contact Technical Support
at
+1-800-635-5577.
216: “LED module
configuration/position
has changes. Please re-
calibrate.”
LED module position
changed from where it
was flat-field
calibrated.
Calibrate instrument for
current LED
configuration.
Place LEDs in previous
calibrated position.
303: “LED modules not
calibrated.”
None of the installed
LED banks have been
flat-field calibrated.
Calibrate instrument with
installed LED banks.
A LED inside the LED
bank burned out.
Calibrate instrument with
installed LED banks.
Replace LED bank if you
are unable to recalibrate
LED bank.
Contact Technical Support
at
+1-800-635-5577.
306: “Flat field calibration
failed. Please verify that
the correct plate is
loaded. Please repeat flat
field calibration.”
Flat Field calibration
failed; see appended
string on the error
message for details.
Calibrate using the
correct Flat Field
calibration plate in the
Read position.
Contact Technical Support
at
+1-800-635-5577.
Table 9-5
Troubleshooting the Optics (cont’d)
Symptom or Error
Message
Possible Causes
Solutions
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...