Troubleshooting
218
0112-0109 H
Troubleshooting the Cell Reservoir
Troubleshooting Data
Table 9-8
Troubleshooting the Cell Reservoir
Symptom or Error
Message
Possible Causes
Solutions
255 “CellReservoir: Reservoir
not detected. Please install it at
Source 3 position and select
Refresh Configuration
from
Instrument
menu bar.”
Instrument cannot
detect the cell
reservoir.
Check to make sure
the Cell Reservoir is
installed.
Check the connection
of the reservoir
sensor inside the
instrument.
256 “CellReservoir: prime sensor
dry during prime/fill. Please
check Fluid Bottle 1”.
Fluid sensor is not
detecting fluid when
pumping from Fluid
1.
Check to make sure
that Fluid 1 bottle is
not empty.
257 “CellReservoir: prime sensor
dry during prime/fill. Please
check Fluid Bottle 2”.
Fluid sensor is not
detecting fluid when
pumping from Fluid
2.
Check to make sure
that Fluid 2 bottle is
not empty.
258 “CellReservoir: prime sensor
dry during prime/fill. Please
check Fluid Bottle 3”.
Fluid sensor is not
detecting fluid when
pumping from Fluid
3.
Check to make sure
that Fluid 3 bottle is
not empty.
259 “CellReservoir: prime sensor
dry during prime/fill. Please
check Fluid Bottle 4”.
Fluid sensor is not
detecting fluid when
pumping from Fluid
4.
Check to make sure
that Fluid 4 bottle is
not empty.
Table 9-9
Troubleshooting Data
Symptom or Error
Message
Possible Causes
Solutions
Negative control wells
show a response.
Wash buffer
components are
different from the
compound buffer.
Make sure wash buffer
and compound buffer are
the same. Check for a
DMSO stimulation, if
present.
Statistical results do not
conform to the expected
results.
Data analysis has not
been optimally set up.
Refer to
.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...