Robotic Integration
244
0112-0109 H
Pause in Well Edit box
The
Pause in Well
edit box on the
Edit Dispense
screen allows the
user to require that the pipettor remain in the well after dispensing.
This pipettor operation will leave the tips in the well for the time
specified. When the tips are in the well, it may not be possible to
exchange plates from any position—and certainly not from the read or
source 3 positions. For this reason, use of this option is not
recommended in remote mode.
Wash Tips Processes
No special requirements are needed for the
Wash Tips
process.
Note that the pipettor and the plate handler do not share the tip
reservoir position. For this reason, plate handling can proceed
unhindered during wash operations.
Mix Fluid Processes
The
Mix Fluids
process does not require any special changes during
protocol creation.
Note that the same warning that applies to
Pause in Well
for
dispensing also applies here.
Also, mixing in any well may interfere with plate handling, so use of this
function may require some experimentation.
Read Processes
No special requirements are needed for
Read
processes.
Finish with Source Processes
This notifies the plate handler to remove the plates in remote mode.
Wash Cell Reservoir Processes
No special requirements are need for the
Wash Cell Reservoir
process.
Command Set
Command Syntax
All commands and parameters should be sent in lower case and end
with <CR>.
All responses are given in lower case and end with <CR>.
Response from ScreenWorks Software has one of the following forms:
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...