data file name containing a full path.
If the data file name parameter is not given, the data file created will
be saved with auto save name as indicated in the
Settings
screen in
the ScreenWorks Software.
Example Command Line
runexperiment<TAB>mydata10312003.fmd<CR>
Example Response
ok<CR> or c50, error string <CR>
Suggested Steps in Command Usage
•
Check if instrument is functioning fine.
•
Make sure experiment is not running.
•
Open protocol first by sending
openprotocol
command.
•
Send
runexperiment
command.
•
Response will be OK, that confirming command started
execution or Condition code related to command, that
confirming command could not start execution.
•
Monitor status for completion, and error status for any other
kind of error.
Command Specific Error Responses
•
Instrument is not functionalCondition code is C00
•
Experiment is runningCondition code is C25
•
Current document is not valid protocolCondition code is C50
•
Invalid data file nameCondition code is C47
Stopexperiment<CR>
Command Description
This command will stop the current action being performed by the
instrument.
The
stopexperiment
command will stop the experiment immediately.
This may halt the instrument pipettor in the midst of a motion. In this
case, the instrument will immediately enter a fault state and an
instrument reset will be required. In addition, the pipettor may not be
in motion but may have tips that contain fluid or are in a source or
target plate. In these cases, it may be possible to recover without
resetting by washing the tips. Regardless, an instrument reset is
recommended following the use of any
stopexperiment
command.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...