Robotic Integration
254
0112-0109 H
Command Parameters
There are no parameters associated with this command.
Example Command Line
stopexperiment<CR>
Example Response
ok<CR> or c55, error string <CR>
Suggested Steps in Command Usage
•
Check if instrument is functioning fine.
•
Make sure experiment is running.
•
Send
stopexperiment
command.
•
Response will be OK, confirming that command execution has
begun, or a Condition code related to the command indicating
that execution could not begin.
•
Monitor status for completion, and error status for any other
kind of error.
Command Specific Error Responses
•
Instrument is not functionalCondition code is 00
•
No experiment is runningCondition code is 55
Clearerror<CR>
Command Description
Cleans error in case recoverable error occurred.
Command Parameters
There are no parameters associated with this command.
Example Command Line
clearerror<CR>
Example Response
ok<CR>
Suggested Steps in Command Usage
•
Check instrument status.
•
Check if error is occurred.
•
Send
clearerror
command.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...