Bug Fix:
Commands made to the remote interface could sometimes take a long
time and cause the remote interface to not respond to the client
application within an acceptable timeframe. Client applications were
able to work around this bug with massive TCP/IP timeouts (minutes)
which made a true hang or network dropout undetectable. This
workaround is no longer required.
The interface itself is unchanged, and if the client is correctly parsing
the ‘status’ response, no changes should be required in the client
application—with the exception of reducing the TCP/IP timeout back to
an acceptable value (for example
,
< 10 seconds). As always, rigorous
regression testing of client code is encouraged.
V1.3
Added new commands to support new functionalities in tetra plus. Four
new commands are
statusex
,
configuration
,
cellflaskcontrol
, and
washcellreservoir
. Noted Tips location has changed from Source 3
location to Source 1 location.
Appendix B—Bar Code Specifications
About Bar Codes
The FLIPR
®
Tetra System bar code scanner is an OEM module with
decode capability to most commonly used bar code symbols in industry
worldwide. The list includes:
•
WPC (UPC /EAN/JAN
•
IATA
•
Code 39
•
Interleaved 2 of 5
•
Industrial 2 of 5
•
Code 128
•
Code 93
•
MSI/Plessey
•
Codabar (NW-7)
Bar Code Recommendations
The bar code reader implementation allows a bar code read down to 7.5
mil (0.20 mm) on photographic-quality symbols and ambient light <
450 ft-candles.
We recommend, however, the user stay above 7.5 mil (0.26 mm) since
most bar code printers may not have the resolution needed to match
the required aspect ratio for a minimum of 3 mm-high symbol lines.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...