0112-0109 H
41
3
Startup and Shutdown
This chapter provides procedures for starting up and shutting down the
FLIPR
®
Tetra High Throughput Cellular Screening System. These
procedures should be followed closely in order to ensure proper
communication between
ScreenWorks
®
Software
and the hardware.
Starting Up the System
To start the FLIPR
®
Tetra System:
1.
Turn on the computer and monitor.
2.
Simultaneously press the
Ctrl+Alt+Delete
keys to launch the
Windows operating system.
3.
At the prompt enter your password.
4.
Turn on the external chiller with the switch located on the left
side of the chiller.
5.
Turn on the FLIPR
®
Tetra System’s power switch located on the
right side of the instrument. The system goes through an
initialization cycle to register all instrument components. This
cycle is not complete until the green
Assay Finished (Unlock)
light is the only light illuminated on the instrument status panel.
6.
Launch the ScreenWorks Software by double-clicking on the
desktop icon.
Note:
After installation, the default password is
flipr
.
Wait for the operating system to finish starting-up before
proceeding.
Note:
Launching the software may take several seconds. Do not
repeatedly double-click the software icon.
Note:
The system is ready for use when the camera temperature in
the instrument status window displays -60 ± 2 °C for the Standard
EMCCD camera, or -20 ± 5 °C for the Aequorin ICCD camera.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...