ScreenWorks Software Overview
116
0112-0109 H
Dispense Configuration (Pin Tool)
The table below describes dispense configuration options when a Pin
Tool head is selected.
Removal Speed
Enter the speed (in mm/s) at which the pin tool is
removed from the source plate. The speed of removal
determines the volume of fluid remaining on the pins.
Note:
Increasing the removal speed increases
the volume of compound on the pin after
aspiration. To decrease the fluid volume, slow
down the removal speed. It is recommended that
assay development be done to determine the
volume transferred based on removal speed.
Pause in Well
Enter a duration (in seconds) to hold the pin tool in the
source plate after full insertion and before removal.
Item
Description
Target Plate
Select the plate to receive the fluid dispense; typically,
the Read Plate.
Height
Enter the distance (measured in
μ
L) from the bottom of
the well the tips should be inserted. The distance
equivalent in millimeters is displayed to the right of the
μ
L value.
Quadrant
Select the quadrant of the target plate into which
compound is to be dispensed.
Note:
Quadrant selection is only available when
liquid is aspirated from a plate using a pipettor
head of lower density, for example, 1536-well
plate using a 384 pin tool head.
Down Speed
Enter the speed (in mm/s) at which the pin tool is
inserted into the target plate.
Removal Speed Enter the speed (in mm/s) at which the pin tool is
removed from the target plate. The speed of removal
determines the volume of fluid remaining on the pins.
Pause in Well
Enter a duration (in seconds) to hold the pin tool in the
dispense plate after full insertion and before removal.
Item
Description
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...