ScreenWorks Software Overview
122
0112-0109 H
Read with TF
Plate reading concurrent with fluid transfer is necessary to collect data
for kinetic cell-based assays as the response time is very rapid. To link
a Read process to a
Transfer Fluid
process, select
Read
in the
Transfer Fluid
page. Inserting a read this way turns the
Read
process
icon purple, indicating that the read occurs concurrently with liquid
addition.
Read with TF
configuration is similar to that of an independent Read
process, but with the addition of two more options;
Number of Reads
Before Dispense
and
Save Images
.
Second Interval Configure the second series of reads taken during the
Read
process. If you do not want a second series with a different
read interval, make
Number of Reads
zero.
Read Interval
—Enter time (in seconds) between reads. The
same interval is used for each read mode (
for example
,
the
instrument takes images twice as fast if a second read mode
is enabled).
Note:
Time should be no less than exposure time
plus read-out time (see
for more information). If it is less, then the
experiment speed will be determined by the capability
of the computer.
Number of Reads
—Enter the total number of reads to be
taken in the second interval.
Item
Description
Note: To remove a
Read with TF
process, uncheck the
Read
option
in the
Transfer Fluid
process dialog box.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...