ScreenWorks Software Overview
124
0112-0109 H
Wash Tips (Standard Pipettor)
The table below describes the options available in the
Wash Tips
process page, for a standard pipettor:
Item
Description
Fluid Type
Select the fluid in the wash reservoir during tip washing,
from wash fluid container A or B.
Pump Speed
Select the speed (slow or fast) you want the pump to fill
and empty the wash reservoir.
Note:
Molecular Devices recommends you use
Fast
except with volatile solvents.
Wash Cycles
From the list, select the number of times that fluid cycles
through the wash reservoir. For each cycle, the pipettor
executes the number in the
Strokes
field.
Strokes
From the list, select the number of times to aspirate and
dispense wash fluid.
Volume/Stroke Enter
the
volume
(in
μ
L) of fluid to be aspirated and then
expelled for each stroke.
Hold Time
Enter the length of time (in seconds) to pause between
aspirating and expelling fluid.
Aspirate Speed
From the list, select the speed (in
μ
L/s) to aspirate fluid.
Dispense Speed
From the list, select the speed (in
μ
L/s) to dispense fluid.
Unload Tips After
Washing
Check this option to have tips removed from the pipettor
head after washing is complete.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...